![Keysight U3606A Скачать руководство пользователя страница 72](http://html.mh-extra.com/html/keysight/u3606a/u3606a_users-and-service-manual_1961516072.webp)
2
Digital Multimeter Operation
Making Measurements
50
U3606A User’s and Service Guide
Testing continuity
1
Press
again (until the continuity annunciator is shown on the
display) to make continuity measurements.
2
Connect the red and black test leads to input terminals
(red) and
LO
(black) respectively as shown in
.
3
Probe the test points. The instrument beeps when the continuity
measurement is less than or equal to the continuity threshold.
Table 2-10
Continuity function summary
Item
Description
Measurement method
0.83 mA ± 0.2% constant current source, open-circuit voltage limited
to < 5 V
Audible tone:
Continuous beep when reading is less than threshold resistance of
10
Ω
at 1.0 k
Ω
range
Input protection
1000 V
rms
on all ranges, < 0.3 A short circuit
Continuity annunciator
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...