Characteristics and Specifications
8
Digital Multimeter Specifications
U3606A User’s and Service Guide
247
AC specifications
Table 8-2
AC accuracy specifications ± (% of r % of range)
Function
Range
[1]
Frequency range
24 Hour
[2]
23 ºC ± 1 ºC
90 Days
23 ºC ± 5 ºC
1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
True rms AC
voltage
[3]
100.000 mV
to 750.00 V
[4]
20 Hz to 45 Hz
0.60 + 0.1
0.60 + 0.1
1.00 + 0.1
[5]
0.02 + 0.02
45 Hz to10 kHz
0.16 + 0.1
0.16 + 0.1
0.20 + 0.1
0.02 + 0.02
10 kHz to 30 kHz
0.80 + 0.1
0.80 + 0.1
1.00 + 0.1
[6]
0.02 + 0.02
30 kHz to 100 kHz
[7]
3.00 + 0.2
3.00 + 0.2
3.00 + 0.2
[8][9]
0.05 + 0.02
True rms AC
current
10.0000 mA
to 3.0000 A
[10]
20 Hz to 45 Hz
0.80 + 0.1
0.80 + 0.1
1.50 + 0.1
0.02 + 0.02
45 Hz to 1 kHz
0.40 + 0.1
0.40 + 0.1
0.50 + 0.1
0.02 + 0.02
1 kHz to 10 kHz
[11]
2.00 + 0.2
2.00 + 0.2
2.00 + 0.2
0.02 + 0.02
[1] 20% over range on all ranges, except for 750 V
ac
range.
[2] Relative to calibration standards.
[3] Specifications stated are for input signals greater than 5% of range.
[4] Available ranges: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V
[5] For 750 V range accuracy is specified for input less than 200 V
rms
.
[6] For 100 mV range, the 10 kHz to 30 kHz accuracy is 1.5 + 0.3
[7] Additional error 0.003% of full scale per kHz to be added when signal input changes less than 10% of range.
[8] For 100 mV range, the 30 kHz to 100 kHz accuracy is 5 + 0.3
[9] For 750 V range accuracy is specified for input less than 300 V
rms
.
[10] Available ranges: 10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A
[11] For 1 A and 3 A ranges, the accuracy is specified for frequencies less than 5 kHz.
N O T E
The specification of the AC+DC measurement will be the sum of the AC and DC accuracy.
The frequency range will be from 50 Hz for 5½ digit resolution and 225 Hz for 4½ digit
resolution.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...