6
Calibration Procedures
Calibration Overview
178
U3606A User’s and Service Guide
Calibration Overview
This section contains procedures for verifying the instrument performance,
as well as procedures for making adjustments where necessary.
Closed-case electronic calibration
The instrument features closed- case electronic calibration. No internal
mechanical adjustments are required. The instrument calculates correction
factors based upon the input reference value you set. The new correction
factors are stored in the nonvolatile memory until the next calibration
adjustment is performed. The nonvolatile EEPROM calibration memory
does not change when power has been turned off, after a Factory Reset
(
*RST
command), or after an Instrument Preset (
SYSTem:PRESet
command).
Keysight Technologies calibration services
Keysight Technologies offers calibration services at competitive prices.
When your instrument is due for calibration, contact your local Keysight
Service Center for recalibration. See
page 239 for information on contacting Keysight.
N O T E
Ensure that you have read the
“Test Considerations” on page 145
before
calibrating the instrument.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...