4
System Related Operation
Utility Menu Summary
124
U3606A User’s and Service Guide
4
Press
Voltage
or
Current
if you wish to configure the ramp signal
parameters for CV or CC output. (The CV or CC annunciator
illuminates respectively when selected.)
5
Select an appropriate output range by pressing
Shift
>
Range
. The ramp
signal amplitude end position is limited by the range selected. (The S1
or S2 annunciator illuminates respectively when selected.)
6
Press
or
to change the ramp signal amplitude end position. See
7
Press
or
until the cursor is position on the ramp signal number
of steps. Press
or
to change the number of steps for the ramp
signal to increment from zero to the amplitude end position. See
Amplitude end position
(NN.NNN V or N.NNNN A)
Number of steps
(NNNNN)
Ramp signal range
CV/CC ramp signal parameters
N O T E
•
The increment of each step in the ramp signal will be the amplitude end
position divided by the number of steps. For example, a 15 V amplitude
end position divided by 100 steps gives an increment of 0.15 V per step.
•
The ramp dwelling time selected will be fastest of the instrument
output capability. (Typically ~300 ms per step.)
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...