DC Power Supply Operation
3
Square-wave Output
U3606A User’s and Service Guide
91
Setting the duty cycle
1
To set the square- wave duty cycle, press
until the duty cycle
value is displayed in the lower secondary display. Ensure that the
square- wave annunciator (
) on the front panel is flashing. (Press
again if it is not.)
2
Press
or
to step through the available duty cycle values.
The duty cycle can be stepped through 256 steps, where each step is
0.390625% more than the previous step. The best resolution the display
can offer is 0.001%.
3
Press
again
or
Shift
>
Save
to save the changes.
Square-wave duty cycle
N O T E
•
Pressing
again will save the changes, and display the next
square-wave parameter. (While still in the edit mode.)
•
Pressing
Shift
>
Save
will save the changes, and exit the edit mode.
•
Changing the square-wave duty cycle value will affect the square-wave
frequency and pulse width values as they are inter-related.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...