Calibration Procedures
6
Adjustments procedures
U3606A User’s and Service Guide
215
9
Verify the constant voltage adjustments using the
readback accuracy” on page 162
CC output adjustment procedure — rear output terminals
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
1
Turn off the instrument and connect the rear panel
+
and
–
output terminals to the
I
(red) and
LO
(black) input terminals.
Table 6-12
Constant voltage output adjustment
Range
Adjustment item
Output voltage
S2 (8 V/3 A)
OUt-L
0 V
OUt-H
8 V
S1 (30 V/1 A)
OUt-L
0 V
OUt-H
30 V
LOAd
[1]
for information on how to connect the LOAd adjustment item.
30 V with additional
30
Ω
, 50 W load
N O T E
Ensure that the rear panel sense terminals (+S and –S) are shorted to the
rear panel output (+ and –) terminals. See
for more information on how to connect the
load leads.