BOP HIPWR 111315
1-15
1.6
FEATURES
1.6.1
DIGITAL CALIBRATION
The BOP Power Supply contains no internal adjustments. Calibration is done entirely via the
keypad (or remotely via the GPIB or RS 232 interface) using digital entries and a calibrated
DVM, a precision d-c reference voltage source and precision shunt resistor. Calibration instruc-
tions appear on the front panel after a password is entered; previous calibration values are
saved and can be restored using the remote interface if desired. The original factory calibration
values can also be restored using the remote interface. (Refer to Section 4.)
1.6.2
VOLTAGE/CURRENT PROTECTION
Positive and negative voltage and current protection values can be individually programmed.
Refer to PAR. 3.3.3.
1.6.3
WAVEFORMS
The BOP models have the capability to make the output follow complex waveforms. These may
be generated either externally, using an analog reference voltage (see PAR. 1.6.6), or internally,
using user-generated waveforms.
Up to 16 user-generated waveforms can be stored for later use. The user selects the operating
mode, positive and negative protection levels, and a count (the number of times the waveform is
to be repeated). Each waveform consists of up to 10 segments; each segment is assigned one
of six basic waveform types: square, sine, triangle, negative ramp, positive ramp and level.
Depending on the waveform type, frequency, peak-to-peak amplitude, offset, and start and stop
phase angle are then assigned by the user.
Segments at the beginning of the waveform can be programmed to be executed only once, or to
be repeated as specified by the count. A graphic representation on the LCD shows the wave-
form as segments are added and is displayed prior to execution. This feature provides unique
versatility for generating a waveform that meets user requirements. Refer to PAR. 3.3.9 for fur-
ther details.
1.6.4
SAVING AND RECALLING SETTINGS
The BOP offers 99 memory locations accessible from the front panel that can be used to store a
set of operating parameters for later use. For each location, the user can store operating mode,
output on/off, Main channel reference type and value, and protection reference type and value.
The stored settings can then be recalled to quickly program the unit to the predetermined set-
ting. Refer to PAR. 3.3.8 for further details.
Содержание BOP-MG 1KW
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 111315 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 39: ...BOP HIPWR 111315 1 19 1 20 Blank FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 40: ......
Страница 56: ...2 16 BOP HIPWR 111315 FIGURE 2 5 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 57: ...BOP HIPWR 111315 2 17 FIGURE 2 6 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 68: ......