3-34
BOP HIPWR 111315
FIGURE 3-11. SAMPLE WAVEFORM
2. The initial waveform settings are displayed (see Table 3-11 and Table 3-12 for parameter
details. Use
!
to modify the initial settings for the waveform. For each parameter, use
$
to save the setting.
Name
CAP CHARG
Mode
VOLTAGE
Protection
Positive
10
Negative
10
Count
0
Segment
Type
Neg Ramp
Frequency (Hz)
50
Amplitude (p-p)
4
Offset
8
When all settings are complete, press
$
to save for power-up. This step creates a repeat-
ing negative ramp starting at +10V decreasing to +6V since it is a 4V p-p signal riding on an
8V offset. The frequency of 50 Hz establishes the time duration of 20 ms for the segment.
3. From the Waveform Settings screen, highlight segment NEG RAMP 50.0HZ 4.00 R and
press
@
to insert a new segment. Use
!
to modify the following settings of the new seg-
ment. For each parameter, use
$
to save the setting.
Type
Level
Period (Sec)
0.002
Offset
0
Initial/Repeat
Initial
When complete, press
$
to save for power-up. This step changes the 2ms, 0V level to be
executed only once at the beginning of the waveform.
4. From the Waveform Settings screen, highlight segment LEVEL 0.002s 0.00 I and press
@
to insert a new segment. Use
!
to modify the following settings of the new segment. For
each parameter, use
$
to save the setting.
Type
Level
Содержание BOP-MG 1KW
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 111315 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 39: ...BOP HIPWR 111315 1 19 1 20 Blank FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 40: ......
Страница 56: ...2 16 BOP HIPWR 111315 FIGURE 2 5 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 57: ...BOP HIPWR 111315 2 17 FIGURE 2 6 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 68: ......