SECTION 4 - CALIBRATION
4.1
GENERAL
This section contains the calibration instructions for the Power Supply. It is recommended that
the user be familiar with Local Mode operation (PAR.3.2) before calibrating the unit.
A full calibration consist of a voltage calibration and a current calibration. Both voltage and cur-
rent calibrations consist of zero, max and min, and protection limit calibration. In addition, the
control signals required for series and parallel operation of multiple units are also calibrated. It is
recommended that the unit be recalibrated annually.
In order to enter the calibration mode the correct password may be required (see PAR. 3.2.4.3).
If the password has been lost, call the factory and a special password will be provided. During
the calibration, new calibration data is computed which is then stored in the nonvolatile memory.
Calibration of the BOP is performed in one of three ways: locally using the front panel keys
(PAR. 4.4) or remotely by sending commands through either the GPIB or RS 232 bus (PAR.
4.3). These two ways cannot be combined.
NOTE: During calibration, various circuits of the BOP are verified. If an error occurs during the
verification process, the unit will beep and an error 24(x), Hardware, (see PAR. B.133)
will be placed in the error queue.
TABLE 4-1. CALIBRATION SUMMARY
Type
Step
Reference Type and
Value
Monitored Parameter
Output Conditions
Notes
VOLTAGE
INTERNAL
1. ZERO
2. POSITIVE
3. NEGATIVE
Internal
Output Voltage
1. 0.0v
2. +Voltage FS
3. –Voltage FS
NO LOAD
AUTOMATIC VM
1. DVM between OUT S
and COM S.
2. Adjust for zero, posi-
tive and negative FS.
VOLTAGE
PROTECTION
1. POSITIVE
2. NEGATIVE
Internal
Output Voltage
1. +Voltage FS
2. –Voltage FS
NO LOAD
AUTOMATIC CM
1. DVM between OUT S
and COM S.
2. Adjust for positive and
negative FS.
VOLTAGE
EXTERNAL
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. +10.0V
(Analog I/O Port,
pin 11 ref to pin 10)
Output Voltage
1. 0.0v
2. +Voltage FS
NO LOAD
AUTOMATIC VM
1. DVM between OUT S
and COM S.
2. Adjust for zero, posi-
tive FS.
VOLTAGE
GAIN
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. +10.0V
(Analog I/O Port pin
11 ref to pin 10)
Output Voltage:
1. 0.0V
2. 0.0V
No Load
Automatic VM
1. DVM between OUT S
and COM S
2. Adjust for zero
(both Zero and Positive).
EXTERNAL
CONTROLS,
SERIES
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. -10.0V
(Par/Ser CTRL-IN
Port, pin 8 ref to
COM S)
Output Voltage:
1. 0.0V
2. +10.0V
No Load
Automatic VM
1. DVM between OUT S
and COM S
2. Adjust for zero,
+10.0V
3. Calibrate unit for
Slave in Series
Содержание BOP-MG 1KW
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 111315 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 39: ...BOP HIPWR 111315 1 19 1 20 Blank FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 40: ......
Страница 56: ...2 16 BOP HIPWR 111315 FIGURE 2 5 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 57: ...BOP HIPWR 111315 2 17 FIGURE 2 6 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 68: ......