BOP HIPWR 111315
3-13
a.
Use the
ADJUST
control to increase or decrease the main channel setting (e.g., volt-
age when the unit is in voltage mode). Start with the most significant digit of the desired
value, then use
T
to highlight the next digit. For fine adjustment press the
ADJUST
control in while rotating the knob to modify the least significant digit.
b.
Enter the desired value on the keypad using the number keys. For example, to program
the BOP to 75.8V, press the following keys in order
75.8
then press
ENTER
. For
fine adjustment use
Y
, and
U
to modify the least significant digit. To correct the entry
before activation press the
CLEAR
key to set the value to zero and start over. When the
desired value is displayed, press
ENTER
. This causes the new value to appear at the
output and be applied to the load if the output is enabled.
TABLE 3-4. VOLTAGE AND CURRENT PARAMETER DEFINITIONS
Term
Definition
To modify
refer to PAR.
Local
Remote
+E
Onom
–E
Onom
The nominal (rated) output voltage of the unit determined by model; e.g. for
a BOP 36-28MG, ±E
Onom
is 36V.
N/A
N/A
+I
Onom
–I
Onom
The nominal (rated) output current of the unit determined by model; e.g. for
a BOP 36-28MG, ±I
Onom
is 28A.
N/A
N/A
+Voltage
–Voltage
Voltage mode only.
Positive (+) and negative (–) output voltage values
established by keypad or remote command.
Range (+): 0 to +Voltage max
Range (–): 0 to –Voltage min
B.87
+Voltage max
–Voltage min
Voltage mode only.
Maximum (positive) and minimum (maximum negative)
voltage that can be set.
Value (+): 0 to +E
Onom
Value (–): 0 to –E
Onom
3.3.4.1
B.89, B.90,
B.91, B.93
+Current Protect
–Current Protect
Voltage mode only.
Defines maximum (+) current and Minimum (maximum
negative) (–) that unit can source or sink.
Range (+): +Current Protect min to +Current Protect max
Range (–): –Current Protect max to –Current Protect min
B.29,
B.30,
B.35, B.33
+Current Protect Max
–Current Protect Min
Voltage mode only.
Maximum setting for +Current Protect and Minimum
(maximum negative) setting for –Current Protect.
Value (+): +Current Protect min to (1.01 x +Current max)
Value (–): –Current Protect max to (1.01 x –Current min)
3.3.4.2
B.37, B.38,
B.39, B.40
Minimum (box)
+Current Protect Min
–Current Protect Max
Voltage mode only.
Minimum (positive) setting for +Current Protect and
maximum (maximum negative) setting for –Current Protect. Values of ±Cur-
rent Protect bCurrent Protect Min and –Current Protect Max (near
zero) are not allowed.This zone (also referred to as the
minimum (box)
is
automatically calculated by the BOP (see Figure 1-3).
N/A
N/A
+Voltage Protect
–Voltage Protect
Current mode only.
Maximum positive (+) and minimum (maximum
negative) (–) voltage that can appear at the output.
Range (+): +Voltage Protect min to +Voltage Protect max
Range (–): –Voltage Protect max to –Voltage Protect min
B.97, B.99,
B.103,
B.101
+Voltage Protect Max
–Voltage Protect Min
Current mode only.
Maximum (positive) setting for +Voltage Protect and
Minimum (maximum negative) setting for –Voltage Protect.
Value (+): +Voltage Protect min to (1.01 x +Voltage max)
Value (–): –Voltage Protect max to (1.01 x –Voltage min)
3.3.4.2
B.105,
B.106,
B.109,
B.107
Содержание BOP-MG 1KW
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 111315 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 39: ...BOP HIPWR 111315 1 19 1 20 Blank FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 40: ......
Страница 56: ...2 16 BOP HIPWR 111315 FIGURE 2 5 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 57: ...BOP HIPWR 111315 2 17 FIGURE 2 6 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 68: ......