B-22
BOP-1K 111315
B.73
[SOURce:]LIST:SET:TRIGger? QUERY
LIST:SET:TRIG?
Syntax:
Short Form: LIST:SET:TRIG?
Long Form: LIST:SET:TRIGger?
Returns <value> = trigger pulse-width between 0.00025 and 0.034 Second
Description
: Returns trigger pulse-width.
If trigger pulse duration has not been set, error 100- “command error”
results
.
B.74
[SOURce:]LIST:SET:WAIT COMMAND
LIST:SET:WAIT
Syntax:
Short Form: LIST:SET:WAIT value
Long Form: LIST:SET:WAIT value
where:
<value> = time-out for a wait command between 0.00025 and 0.034 Seconds
Description:
Establishes the time-out for a wait command.
If this command is not sent, the time out period will
be 0 indicating an infinite wait time. (See Figures B-6 and B-7.)
B.75
[SOURce:]LIST:SET:WAIT? QUERY
LIST:SET:WAIT?
Syntax:
Short Form: LIST:SET:WAIT?
Long Form: LIST:SET:WAIT?
Returns <value> = time-out for a wait command between 0.00025 and 0.033 Second
Description
:
Returns the time-out for a wait command. Returns 0 if LIST:SET:WAIT not sent, indicating infinite wait
time.
B.76
[SOURce:]LIST:TRIGger COMMAND
LIST:TRIG
Syntax:
Short Form: LIST:TRIG <value>
Long Form: LIST:TRIGger <value>
where:
<value> = programmed current (Amps) for current list, or programmed voltage (Volts) for voltage list.
Description:
Similar to LIST:VOLT or LIST:CURR, except also allows Output Off flag of External Protection
Port to be used to trigger an external device.
In addition to setting the output to the <value> speci-
fied (for duration established by LIST:SET:TRIG), causes the Output Off flag transistor (EXT_E, pin 5,
and EXT_C, pin 6, of the External Protection port) to be set to the state (on or off) for the duration set
by the LIST:SET:TRIG command.
If the trigger duration is not established by the LIST:SET:TRIG command, a -221 “Settings conflict”
error results. This command must be preceded by LIST:CURR or LIST:VOLT command, otherwise a -
221 “Settings conflict” error results. The entries on the dwell list must match the values on the current
or voltage list, otherwise a -226 “Lists Not Same Length” error results. (See Figure B-7.)
B.77
[SOURce:]LIST:VOLTage COMMAND
LIST:VOLT
Short Form: LIST:VOLT[:LEV] <exp_value>, <exp_value>, . . . (to max of 5900 data points)
Long Form: LIST:VOLTage[:LEVel] <exp_value>, <exp_value>, . . . (to max of 5900 data points)
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1
Description:
Adds the voltage value (in Amps) to list.
This command sequentially adds LIST:VOLTage values to
the main channel List Data Table locations illustrated in Table B-2. LIST:CLE sets starting location to 0.
Next location indicated by LIST:VOLT:POIN? The maximum number of entries is 2950, 3933, or 5900.
Since the input buffer of the BOP has a limit of 253 characters, multiple commands are necessary to
complete the full 5900 entries of the list. If LIST:CURR has any entries, an error message: -221,”Set-
tings conflict” is posted in the error queue. Related Commands: LIST:VOLT:POIN?, FUNC:MODE,
LIST:CLE, *RST. (See example, Figure B-5.)
Содержание BOP-MG 1KW
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 111315 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 39: ...BOP HIPWR 111315 1 19 1 20 Blank FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 40: ......
Страница 56: ...2 16 BOP HIPWR 111315 FIGURE 2 5 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 57: ...BOP HIPWR 111315 2 17 FIGURE 2 6 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 68: ......