2-10
BOP-1K-GL 111315
FIGURE 2-3. FACTORY DEFAULT POWER-UP SWITCH SETTINGS
2. Connect a twisted wire pair (either #24 or #22 AWG) to the mating connector for the Analog
I/O port pins 11 and 10. C10V d-c ±0.1mV to pin 11, referenced to pin 10, then
install the mating connector on the Analog I/O port at the rear panel.
CAUTION: DO NOT repeatedly toggle the POWER circuit breaker/switch as this may
damage the unit.
3. Connect the power supply to source power (see PAR. 2.5.2). With no load connected, set
POWER switch to the ON position.
4. Each time the unit is turned on an internal self-test is performed. If the unit passes, it initial-
izes with the power-up settings established in step 1 and the front panel POWER/FAULT/
LIMIT light is green (power good), the DIGITAL CONTROL light is not lit (analog input
enabled) and the VOLTAGE/CURRENT light is green (voltage mode), MASTER/SLAVE light
is green (standalone or master configuration) and the OUTPUT ON light is lit (output
enabled). If the front panel POWER/FAULT/LIMIT light is red, the unit has failed self-test;
contact Kepco for further instructions. If the unit beeps, or the MASTER/SLAVE or VOLT-
AGE/CURRENT light blinks, refer to PAR. 3.8 for troubleshooting.
5. Connect a digital voltmeter (DVM) (resolution and accuracy of 0.01% or better) to the OUT S
and COM S terminals at the rear panel terminal block.
6. Verify DVM voltage reading matches the nominal voltage of the unit within 0.01% of rated
maximum voltage.
2.3.2
PRELIMINARY OPERATIONAL CHECK USING DIGITAL CONTROL
1. With POWER switch set to off position, verify that the power-up switches at the top cover are
set to the factory default positions shown in Figure 2-3. This establishes GPIB address 6,
standalone operation, Voltage mode, Analog Input enabled and internal ±current limits
enabled and set to maximum.
2. Set Power-up switch S3-2 to 0 (off) to disable Analog Input.
NOTE: The unit is shipped with load type set to Active (for inductive loads). If necessary, refer
to 3.3.6 for an explanation of load types and to PAR. 3.3.2.1 to change the load type.
3. Connect a twisted wire pair (either #24 or #22 AWG) to the mating connector for the Analog
I/O port pins 11 and 4. C10V d-c ±0.1mV to pin 11, referenced to pin 4., then install
the mating connector on the Analog I/O port at the rear panel.
4. Connect the GPIB or RS 232 cable coming from the host computer to the appropriate port
(GPIB or RS 232) at the BOP rear panel.
S1: 00110 = GPIB ADDRESS 6
S2: 00001 = STANDALONE
S3: 11111 = VOLTAGE MODE,
ANALOG INPUT ON
±V AND ±C PROTECT MAX SET
TO MAXIMUM