LIST:CLEAR (PAR. B.49). Always precede a new list with this command. A list remains in mem-
ory until the power is cycled or the LIST:CLEAR command is processed.
LIST:DWELL (PAR. B.60). Defines the dwell time for each point in a list. In many instances it is
easier to use one dwell time and repeat a specific point multiple times to generate longer dura-
tion pulses. Remember, if more than one LIST:DWELL is sent, there must be a LIST:DWELL for
each voltage or current point in the list.
LIST:COUNT (PAR. ) Defines how many times a list is executed. 0 equals indefinite; when a
count of 0 is used, either *RST, VOLT:MODE FIX or CURR:MODE FIX must be used to stop the
list. *RST will cause the output to be set off and the unit is set to Voltage mode. VOLT:MODE
FIX or CURR:MODE FIX stops the list immediately; the point being executed when the list is
stopped will be present at the BOP output.
3.7.3.3.2 OTHER REQUIRED COMMANDS
VOLT:MODE LIST (PAR. B.94) or CURR:MODE LIST (PAR. B.27). These commands start the
list and VOLT:MODE FIX (PAR. B.94) or CURR:MODE FIX (PAR. B.27) stop the list. LIST com-
mands are not accepted while a list is running. If LIST:COUNT is between 1 and 255, when the
count decrements to 0, the list stops and the unit automatically moves to VOLT:MODE FIXED or
CURR:MODE FIXED state. To execute the list again, either VOLT:MODE LIST or CURR:MODE
LIST must be sent again.
3.7.3.3.3 OTHER USEFUL COMMANDS
LIST:COUNT:Skip (PAR. B.53). This command provides the ability to execute the initial points
only once whenever a list is run. It is used to set initial preconditions prior to running a repetitive
sequence.
LIST:VOLT:POINTS? (PAR. B.79) or LIST:CURR:POINTS? (PAR. B.59) These queries return
the number of points in a list and provide a simple way to insure that all points entered were
properly processed and as intended.
3.7.3.3.4 OPTIONAL COMMANDS
Most commands have associated Queries (?) These are useful for troubleshooting/debugging
lists but are not needed in most cases.
3.7.3.4
MEASURE SUBSYSTEM
This query subsystem returns the voltage and current measured at the power supply's output
terminals.
3.7.3.5
OUTPUT SUBSYSTEM
This subsystem controls whether the power supply output is disabled or enabled.
Содержание BOP-GL 1KW
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 111315 FIGURE 1 1 HIGH POWER BOP GL SERIES POWER SUPPLY...
Страница 36: ...1 16 BOP 1K GL 111315 FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 38: ......
Страница 116: ......
Страница 128: ......
Страница 174: ......