BOP-1K-GL 111315
2-17
2.7.2
SETUP FOR ANALOG CONTROL
1. With power off configure the power-up setup switches (see Table 2-2 for details). Setup for
analog control can be accomplished through the power-up switches in one of two ways:
a. Setting S3-5 (MSB) through S3-1 (LSB) to 0 (off) [00000] is a unique configuration that
allows external signals applied to the I/O port to control the BOP upon power-up: Analog
Input ON (main channel reference, pin 11 of Analog I/O port) enabled, limit channel refer-
ences (pins 5, 6, 13 and 14 of Analog I/O port) enabled, Voltage/Current mode is estab-
lished by pin 2 of Analog I/O port (1 or open = voltage mode, 0 or short = current mode)
and output is set to ON (enabled).
b. Setting S3-5 to 1 allows S3-1 through S3-4 to determine how the unit is controlled.
•
Set S3-1 to either 1 for Voltage mode or 0 for Current mode
•
Set S3-2 to 1 to enable analog control (allow the main channel to be controlled by the
external reference applied to analog I/O Port pin 11).
•
Set S3-3 and S3-4 to configure the protection limits (see Table 2-2).
2. Connect the load and programming and monitoring equipment as shown in Figure 2-6.
2.7.3
SETUP FOR DIGITAL CONTROL VIA GPIB
1. With all power off, connect the load to the BOP using either local or remote sensing. If units
are to be connected in series or parallel, refer to PAR. 2.8. Connect the GPIB connector to
the GPIB port (see Figure 2-1 and Table 2-1).
2. Refer to Table 2-2 and configure GPIB address, power-up settings, and master/slave status.
Set S3-5 to 1 and S3-2 to 0 to set Analog Input to OFF.
3. Refer to PAR. 3.3.2.2 to turn power on.
4. Refer to PAR. 3.3 for power supply basics and PAR. 3.6 for remote mode programming.
SCPI programming is described in PAR. 3.7; Appendices A and B provide syntax for SCPI
common and subsystem commands and queries implemented in this unit.
2.7.4
SETUP FOR DIGITAL CONTROL VIA RS 232C
1. With all power off, connect the load to the BOP using either local or remote sensing (refer to
PAR. 2.5).
2. If units are to be connected in series or parallel, refer to PAR. 2.8.
3. Connect the RS 232 connector to the RS 232 port (see Figure 2-1 and Table 2-1).
4. The default baud rate is 9600. The baud rate can be set to 19.2K by performing reset power-
up (PAR. 3.3.2.1) with switch S3-1 set to 1.
5. Refer to PAR. 3.3.2.2 to turn power on.
Refer to PAR. 3.3 power supply basics and PAR. 3.7 for RS 232 operation. SCPI programming
is described in PAR. 3.7; Appendices A and B provide syntax for SCPI common and subsystem
commands and queries implemented in this unit. Refer to PAR. 3.6.4.3 to configure the RS 232
Port.