7 Cephalometric calibration and alignment
98
To complete this calibration:
1. Select the
CEPH Mech Adj
program.
2. Press
Patient Positioning
.
3. Take an exposure.
4. Follow the instructions in the calibration result image.
Possible instructions and references to further informa-
tion are found in table “instructions for adjustment”.
5. Take new exposures until the calibration result is
“Passed”.
Table 8.1 Instructions for adjustment
Instruction
Additional information
Check for obstacles in the primary collimator. Turn away
the nasion support.
Something is blocking the X-ray
beam. Check for wrong nasion sup-
port position, loose parts in the colli-
mator assembly or calibration tools
attached to the panoramic/3D chin
support.
NOTICE!
After turning the cepha-
lostat, the Ceph Rot Angle calibration
must be redone. The x-ray beam will
otherwise not be central at the cepha-
lostat sensor.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...