8 Special procedures
133
8.7 Firmware update
8.7.1
Preparing for update
You need the following for a typical update:
The latest release package, available from Technical
Support’s extranet site.
A host PC for performing the update. The host PC may be
a separate PC (e.g. a laptop computer) or the image
acquisition workstation.
Also recommended is a spare CAT6 Ethernet cable and a
USB Blaster (see chapter
Volatile installation
).
Preparing the update
1.
Prepare a connection to the X-ray unit.
When performing the update from the image
acquisition workstation, close down any imaging
application.
When performing the update from a separate
computer, connect an Ethernet cable directly between
the computer and the X-ray unit. For this to work, you
need to make the computer use the same subnet as
the X-ray unit. On Windows 7 (similarly for other
Windows versions), select
Control Panel
→
View
Network Connections
→
Local Area Connection
→
Properties
→
Internet Protocol Version 4
and click
the
Properties
button.
Use the following settings:
IP address: 10.208.6.102
Subnet mask: 255.255.255.0
NOTICE! The settings above are valid if the X-ray unit
is using the default IP address (10.208.6.101). If the
unit’s IP address is something else, set the first three
numbers in the computer’s IP address equal to the
unit’s, and make the last one differ. The unit’s IP ad-
dress is shown in the settings view in GUI.
2.
Copy the software release package to the host PC and
unzip it.
3.
Locate openS2.bat in the unzipped release package
and double click it. You will be prompted for unit’s IP ad-
dress. Type it in (if other than default) and press enter.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...