8 Special procedures
142
8.
Mark the Program/Configure check box.
9.
Click the Start button and wait for the download to com-
plete (takes a few seconds).
10. Volatile installation is typically followed by a permanent
install. Refer to chapter Manual update for instructions
on how to do this. The volatile install is voided if the unit
is powered off before the software is permanently in-
stalled.
8.9 Special cases
8.9.1
Programming a R3220 Main CPU
board having wrong unit type
configured
In case the CPU board is configured with wrong unit type
or does not have a unit type configured, the s2terminal will
not be able to locate the programming files properly. This is
possible, for example, if a spare part R3220 board is used.
To decide the unit type assumed by the s2terminal, follow
the messages printed in the prompt right after a connection
has been established.
If the board's unit type is not recognized by the s2terminal,
it will prompt the user for the type to use. Enter the unit's
type and program the board according to the instructions in
chapter 8.7.2.2 Manual update.
If the board's unit type is not what it should be,
programming must be done using an explicitly named
programming file. See instructions below.
To program a R3220 Main CPU board using explicitly
named programming files:
1.
Establish a connection to the board as per instructions
in Preparing for update.
2.
Update the FGPA core using the command:
xcpr 3220 OP3DPRO_R3220_R<ver_ma-
jor>_<ver_minor>.pof
Example:
xcp r3220 OP3DPRO_R3220_R2_00.pof
Wait until the software is transferred and power off the
unit.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...