6 Panoramic and 3D calibrations and adjustments
92
5. Visually evaluate the result using the installed imaging
software.
Subjects to be evaluated:
1. Smoothness of the exposed area. Non-exposed area
surrounds the whole image.
2. High contrast resolution; minimum 3.1LP/mm must be
distinguishable.
3. All four low contrast holes must be visible.
4. Roundness of the ball.
5. The ball should be placed symmetrically between the two
pins. The distance from both pins to center should be
equal length.
Re-do panoramic geometry calibration if above conditions
are not met.
NOTICE! The panoramic QC collimator is equipped with a
0.8 mm copper filter. If more filtration is required, additional
filtration may be attached to the tubehead cover. The unit
may be configured to use higher exposure values to
compensate for an additional 1 mm copper filter using
service command copperthickness. Refer to service
command help copperthickness for further instructions.
1
2
3
4
5
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...