6 Panoramic and 3D calibrations and adjustments
86
The results of the periodical checks below should be
recorded and the check images filed. To do this, it is
recommended that a “Constancy Test” “Patient” record is
created under which the test images can be recorded.
Please note that the reference levels may change with x-
ray unit firmware / driver updates. Upon such updates, new
reference levels for measured values for the unit should be
established. Always refer to the limiting values that are
specified with current 3D QC test.
3D Quality Check program execution
1. Attach the QC phantom to the unit.
2. Select the
3D QC
program.
3. Press
Patient Positioning
.
4. Take an exposure.
NOTICE! Reference your 3D imaging program manual
for instructions on viewing the 3D volume.
3D QC program result evaluation
NOTICE! Reference your 3D imaging program manual for
instructions on viewing the 3D volume.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...