8 Special procedures
149
1.
Wrong command syntax or binary file name typed in.
Check the syntax and filename.
2.
The files are not found in the assumed location. Run the
s2terminal using the openS2.bat file found in the soft-
ware release package's root folder. Check that the file
mentioned in the error message is really present. If not,
unzip a new copy of the release package.
The unit type of the Main CPU board is incorrectly set. The
unit type for which the s2terminal is trying to find binary
files is indicated in the error message. See chapter Special
cases / Programming a R3220 Main CPU board having
wrong unit type configured for instructions on how to solve
this issue.
8.11 Configurable panoramic mAs limit
A mAs (milliampere-seconds) limit of 250 mAs may be set
for panoramic programs, if required by local regulations.
When mAs limiting is enabled, the largest mA setting is
removed for the imaging programs Standard panoramic,
OrthoZone and Wide arch panoramic.
The mAs limiting feature is controlled by the s2terminal
service program
limitmas
:
limitmas
prints current setting
limitmas on
enables limiting
limitmas off
disables limiting
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...