8 Special procedures
130
■
For 220-240Vac use two 10A fuses, type T10A H
250V (Littelfuse 326 010)
■
For 100-120Vac use two 15A fuses, type T15A H
250V (Littelfuse 326 015)
6.
Loosen the screws (DO NOT REMOVE) of the mains
inlet cover and remove the cover.
7.
Install the mains power cord wires to the terminal block
on the base of the column and route the cable through
the strain relief.
■
For EU 230Vac: H05VV-F 3G 1.5mm2 with a
Schuko plug (CEE 7/7)
■
For USA/CAN 115Vac: SJT AWG14 with a hospital
grade NEMA 5-20 plug
■
For USA/CAN 230Vac: SJT AWG14 with a hospital
grade NEMA 6-15 plug
8.
Install all the covers.
WARNING! Before connecting the mains voltage to the
unit, check that the installation environment's tempera-
ture and humidity complies with allowed operating con-
ditions of the unit. Make sure that the power line meets
the requirements set by the manufacturer. See the
Technical Specifications in the User Manual for more
details.
9.
Now the unit can be connected to the mains and be
switched on.
NOTICE! If the unit is moved to a new location, check
that the voltage at the new location is the same as the
unit is configured to. If not, the unit needs to be config-
ured to that particular mains voltage.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...