7 Cephalometric calibration and alignment
111
5. When the adjustments are done press the exposure but-
ton (no X-rays are generated) to acknowledge that the
check has been carried out.
7.2.7
Cephalostat pixel calibration
1. Attach secondary collimator and tubehead covers.
2. Rotate the ear holders into PA view position and move
them completely apart.
NOTICE! Turn nasion support up out of the way.
3. Select
CEPH Pix
Calibration program.
4. Press
Patient Positioning
.
5. Take an exposure.
6. This calibration should always be passed.
7.2.8
Cephalostat Quality Check (optional)
NOTICE! Rotate the ear holders into PA view position.
Turn nasion support up out of the way.
NOTICE! Not required in North America.
1. Attach the QC phantom to the cephalostat unit and en-
sure that it’s leveled from the spirit level.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...