1 Introduction
4
1.6 Warnings and precautions
1.6.1
Warnings for cross infection
1.6.2
Warnings to be observed during
installation and service
Always use available disposable protective covers with the
patient positioning accessories:
• Bite fork cover
• Chin support cover
• Head support cover
• Nose support cover
• Ear holder cover
In case the transportation package has been damaged,
make sure the actual product is fully intact.
Before attempting to service the device make sure that you
know how to operate it. Read the user’s manual.
Read and familiarize yourself with the warnings and precau-
tions listed in the user’s manual.
Only use original spare parts from the manufacturer when
repairing the device or replacing parts.
Warning - Radiation Safety
Before servicing the unit familiarize yourself with local and
national radiation safety standards and requirements relat-
ing to dental x-ray equipment.
Warning - Electrical Safety
Disconnect the unit from the main power supply before re-
moving any covers.
Disconnect the unit from the main power supply before re-
pairing or replacing mechanical parts or installing accesso-
ries.
Be careful when operating the unit not to get body parts or
clothing trapped between moving parts.
Disconnect the unit from the mains power connection be-
fore servicing the unit, e.g. replacing circuit boards or other
electrical components.
If there are capacitors on a circuit board or electrical device
wait ten (10) minutes, after disconnecting the unit from the
power supply, before handling the board or device.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...