8 Special procedures
140
ware drop down list and click Close.
5.
Click the Auto Detect button to have the programmer
detect the device type and verify the connection to the
circuit board.
6.
Select the first row in the list and click the Change
File… button.
7.
Select the correct sof file for the circuit board (e.g.
R3300_R2_00.sof for release R2.00 and R3300 circuit
board).
8.
Mark the Program/Configure check box.
9.
Click the Start button and wait for the download to com-
plete (takes a few seconds).
10. Volatile installation is typically followed by a permanent
install. Refer to chapter Manual update for instructions
on how to do this. The volatile install is voided if the unit
is powered off before the software is permanently in-
stalled.
8.8.1.2
R3220
1.
Connect the USB-Blaster's USB connector to a PC
having the Quartus II Programmer software installed.
2.
Connect the USB-Blaster's 10-pin connector to the
JTAG port on the circuit board to be programmed (see
circuit board specific connections below). Note the ori-
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...