6 Panoramic and 3D calibrations and adjustments
70
6.5.5
Panoramic collimator calibration
NOTICE! With Pan unit (no 3D option) this calibration is
done by adjusting the collimator.
NOTICE! There is a 6-8 second delay after initially
pressing the exposure button before exposure occurs.
1. Attach the panoramic sensor.
2. Select the
PAN Col
program.
3. Press
Patient Positioning
.
4. Take an exposure.
5. The status is displayed on the image captured in the im-
aging software, how much the height needs to be adjust-
ed and in which direction.
6. Adjust the height of the PAN sensor and check the tilt and
position of the collimator. Make adjustments and take
new exposures until calibration result “passed” is
achieved.
First two lines are for DIN and the rest for the IEC standard.
Field of useful beam (h x w): 126.0 x 129.2 mm result according to DIN.
Field of useful beam max. (h x w): 138.7 x 140.3 mm DIN limit.
X-ray field (h x w): 126.7 x 131.0 mm result according to IEC.
X-ray field max. (h x w): 138.7 x 140.3 mm IEC limit.
X-ray field discrepancy (up+down+left+right): 12.9 mm result according to IEC.
X-ray field discrepancy (up+down+left+right): 22.7 mm IEC limit.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...