6 Panoramic and 3D calibrations and adjustments
62
performed as a support for the installer. All calibrations
must be done until the system informs they have passed.
The exclamation marks disappear when a passed
calibration has been saved.
NOTICE! After calibrations place all the needed patient
positioning accessories back to the chin rest.
NOTICE! Press Save on GUI after each calibration.
6.2 When to calibrate the unit
The unit must be calibrated and, if necessary, adjusted at
regular intervals in accordance with national regulations
regarding the use, maintenance and service of dental x-ray
devices that are in force in the country in which the unit is
installed.
In any event, the unit must be completely re-calibrated at
least once a year.
The unit must also be fully or partially calibrated when
parts are replaced. The calibration steps for the following
replaced parts are as follows:
Tubehead
After replacing the tubehead the unit must be completely
recalibrated.
NOTICE! Also perform Quality Check when you replace
the tubehead.
R3800 X-ray generator
After replacing R3800 X-ray generator preheat and mA
calibration must be redone.
Collimator assembly
After replacing or adjusting the collimator assembly redo all
applicable collimator, pixel and geometry calibrations as
well as the quality checks.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...