8 Special procedures
135
6) Update the R3300 firmware and core.
7) Update the Ceph R3210 core (units equipped with
cephalostat).
8) Power off and on again.
9) Update the Ceph firmware (units equipped with
cephalostat).
10) Set software release version.
After updating any software component, the unit must be
restarted before normal operation is resumed. Reboot is
not required between separate software component
updates, unless otherwise stated in the component's
update instructions.
Main CPU FPGA core update
1.
Press the F2 button and press enter, or type in manu-
ally:
xcc r3220 R<ver_major>.<ver_minor>
Example:
xcc r3220 R2.00
This starts a R3220 core update. Wait until the software
is transferred and the unit has restarted. Power off the
unit before the countdown in the boot prompt ends.
2.
Power on the unit again, this time type:
safe
and then press enter before the countdown in the boot
prompt ends. This puts the unit in safe mode for R3220
firmware update.
Main CPU firmware update
Press the F3 button and press enter, or type in manually:
xf R<ver_major>.<ver_minor>
Example:
xf R2.00
Wait for the software package to be transferred and the
unit to restart itself. If the update fails, try again.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...