Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
7-10
Resetting all Filters to the Default Values
To reset filters to the default values:
1. From the Main menu, select
Tools >
Filter Settings > Reset To
Default
.
2. Click
Yes
. The dialog box closes and all filter settings are set to
the default.
Taking Measurements
Hounsfield Units
The system software allows making measurements in Hounsfield
Units (Shape Region) for all image views. These measurements
calculate and display the average (Mean) grayscale level of the area
enclosed from -1000 to 3000 (where 0 equals the density of water).
The Standard Deviation is also calculated, where the smaller the
number, the closer each shade of gray is in density to the others in
the enclosed area.
1. To make a measurement, right click the image and select
HU
Statistics
.
The menu closes and the cursor changes to the
hu
measurement symbol.
2. Click, drag, and click to define an area. Measurement
statistics appear in the upper right corner of the image.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...