k990400 September 19, 2007
General Product Information
12-5
Electromagnetic or other Interference
(Emissions and Immunity)
The System was tested and found to comply with the limits for Class B equipment,
pursuant to IEC 60601-1-2. These limits are designed to provide reasonable
protection against harmful interference in a commercial environment. The system
generates, uses, and can radiate electromagnetic energy and if not installed and used
in accordance with these instructions, may cause harmful interference to
surrounding equipment. Also portable and mobile RF communications equipment
can effect the operation of the system. Proper precautions and equipment location
should be observed, refer to the following table:
The System is intended for use in an electromagnetic environment specified below. The
Customer or User should ensure that the system is used in such an environment.
Immunity Test
Compliance Level
Electrostatic Environment - Guidance
IEC 61000-4-2
Electrostatic
Discharge (ESD)
+6 kV Contact
+8 kV Air
Floors should be wood, concrete or ceramic tile. If
floors are synthetic, the relative humidity should be
at least 30%
EFT
IEC 61000-4-4
+2 kV Mains
+1 kV I/Os
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
+1 kV Differential
+2 kV Common
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips/
Dropout
IEC 61000-4-11
>95% Dip for 0.5 cycle
60% Dip for 5 cycles
30% Dip for 25 cycles
>95% Dip for 5
seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user
requires continued operation during power mains
interruptions, it is recommended that the system be
powered from an uninterruptible power supply.
Power Frequency
50/60 Hz
Magnetic Field
IEC 61000-4-8
3A/m
Power frquency magnetic fields should be that of a
typical commercial or hospital environment.
Conducted RF
IEC 610000-4-6
3 Vms 150 kHz to 80
MHz
Portable and mobile communication equipment
should be separated from the system by no less
than 19 feet (6 meters).
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...