k990400 September 19, 2007
Calibration and Quality Assurance
10-15
10. The white dotted line, in the preview screen, is use to set the
water height. The line should appear near the middle of the water
height.
Adjust the Phantom Platform to achieve the proper height. Use
the menu
Back/Front
arrow-slide to adjust the phantom side-to-
side position.
If required, make adjustments and click Preview again. Repeat
as required.
11. Click
Capture
to start the scan process.
12. Access the HU measurement image for evaluation by double
clicking
Coronal View
(bottom image, 3
rd
from the left).