Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
11-2
Conditions of Operation – 9 and 27 Second Scans
All data was acquired using Victoreen Model 450P Survey Meter (S/N 3159) in
integrate mode. The data was acquired in concentric circles of radii listed in table
below A lab head phantom was placed within the beam to act as the scattering agent.
The Survey Meter was placed at a height of 112 cm [44 in] while the mid-point of
the beam exit window was located 118 cm [46.5 in] above the floor. The detector
was in the portrait position and X-ray field was at its maximum opening.
Locations listed in the table are for 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°.
Looking out from the system 0° was located directly in front (A location).
This table consists of data gathered with the system operating at 120 kVp, 5 mA
with an assumed use and occupancy factor of 1 and the following parameters:
Estimated usage: 10 scans per week yielding a workload of 16.7 mA-min/week
25 scans per week yielding a workload of 41.7 mA-min/week
50 scans per week yielding a workload of 83.3 mA-min/week
Total scan time:
9 seconds
Location
Distance in
Feet
[meter]
Exposure
(mR)
Exposure
(
μ
R)
10
scans/wk
mR/wk
25
scans/wk
mR/wk
50
scans/wk
mR/wk
0
o
(A)
3 [0.91m]
0.120
119.81
1.20
3.00
5.99
6 [1.82m]
0.035
34.94
0.35
0.87
1.75
9 [2.74m]
0.018
18.47
0.18
0.46
0.92
45
o
3 [0.91m]
0.140
139.77
1.40
3.49
6.99
6 [1.82m]
0.041
41.43
0.41
1.04
2.07
9 [2.74m]
0.023
23.46
0.23
0.59
1.17
90
o
(B)
3 [0.91m]
0.140
139.77
1.40
3.49
6.99
6 [1.82m]
0.055
54.91
0.55
1.37
2.75
9 [2.74m]
0.018
17.97
0.18
0.45
0.90
135
o
3 [0.91m]
0.015
15.47
0.15
0.39
0.77
6 [1.82m]
0.010
10.48
0.10
0.26
0.52
9 [2.74m]
0.006
5.99
0.06
0.15
0.30
180
o
(C)
3 [0.91m]
0.062
61.90
0.62
1.55
3.09
6 [1.82m]
0.017
17.47
0.17
0.44
0.87
9 [2.74m]
0.008
8.49
0.08
0.21
0.42
225
o
3 [0.91m]
0.014
13.98
0.14
0.35
0.70
6 [1.82m]
0.006
6.49
0.06
0.16
0.32
9 [2.74m]
0.003
3.49
0.03
0.09
0.17
270
o
(D)
3 [0.91m]
0.140
139.77
1.40
3.49
6.99
6 [1.82m]
0.038
37.94
0.38
0.95
1.90
9 [2.74m]
0.014
14.48
0.14
0.36
0.72
315
o
3 [0.91m]
0.110
109.82
1.10
2.75
5.49
6 [1.82m]
0.030
30.45
0.30
0.75
1.52
9 [2.74m]
0.013
13.48
0.13
0.34
0.67
1 meter above
0.004
4.49
0.04
0.11
0.22
1 meter above and 1
0.011
10.98
0.11
0.27
0.55
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...