2. Prepare the Bite Tip by inserting
the narrow edges of white Bite
Tip down into Bite Tip Holder
uprights. Then turn the Bite Tip
a 1/4 turn to lock into place.
The Patient will be biting on the
grove of the white Bite Tip
during the scan.
3. Insert the Chin Rest and Bite Tip
Holder into the Positioning Block as shown below.
4. Adjust Patient height to where the Patient’s chin is on the Chin
Rest, but keeping the Erect posture.
NOTE:
The front-to-back Patient positioning is critical for a quality
image. The PAN scan capture is a tall but narrow window.
5. Patient must bite in the grove of the white bite piece to ensure
proper front-to-back positioning.
Narrow
Edges
Patient Tilted
Downward
10
o
CHIN REST
(Clear or White in Color)
Bite Tip
Holder
Horizontal Laser
Occlusal P
lane
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...