k990400 September 19, 2007
12-1
Chapter
12
General Product
Information
Technical Specifications
X-ray Source:
Tube Voltage:
120 kVp
Tube Current:
3-7 mA
Voltage Wave Shape: Constant Potential
Focal Spot:
0.0197 inches (0.5 mm)
Duty Cycle:
3%
Source to Sensor distance:
28.1 inches (71.4 cm)
Source to Patient distance*:
19.5 inches (49.53 cm) (center of rotation)
* The patient must be properly positioned in the Head Support
Positioner Mechanism for each patient for all applications in
order to have the focal spot to skin distance as large as possible.
Minimum Filtration (at 120 kVp) (mm of aluminum equivalent):
10 mm or greater
Maximum Rated Continuous Tube Operation:
130 kVp @ 0.5 mA
Maximum Rated Pulsed Tube Operation:
130 kVp @ 1mA
NOTE:
Leakage technique factors are measured at the maximum
specified energy.
Maximum Deviation:
kVp: +/- 5 kVp
mA:
+ 10%
Timer:
±
.01 seconds or 5%, whichever is greater
X-ray Beam Size:
Rectangular cone 9.37” wide x 1.97” to 7.56” high
(23.8 cm width x 5 cm to 19.2 cm height)
i-PAN Option:
Rectangular cone 0.39” width x 6.30” height
(1cm width x 16 cm height)
(Automatically collimated not to exceed image detector readable area)
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...