background image

Kavo 3D eXam ®  Operators’ Manual

k990400    September 19, 2007

9-6

3. Choose a path for the 

Export DICOM Destination Folder

.

If required, click 

Make New Folder

.

4. Click 

OK

.

Reporting

Patient reports are generated by inserting patient images into a single 
or multiple page document. Patient information and text notes can 
also be inserted as required. Once these reports are created and 
saved, they are readily available for printing and distribution. Patient 
reports can also be accessed for editing.

From the top menu bar, select 

Tools > Reporting

 to display the 

following three report options:

Run Report

Create New Report

Modify Existing Report

Exit

Run Report

This feature enables the user to access a sample report to modify and 
and save as a new patient report.

Содержание 3D eXam

Страница 1: ...k990400 September 19 2007 Cone Beam Volumetric Tomography and Panoramic Dental Imaging System Operators Manual...

Страница 2: ...iences International please write call or fax to Imaging Sciences International at the above address This documentation has been drafted approved supplied in the English Language Imaging Sciences Inte...

Страница 3: ...ngs Cautions and Notes 2 1 Safety Precautions 2 2 Electrical Fire Hazard 2 2 Mechanical Hazards 2 3 Laser Beam Hazards 2 3 Radiation Safety 2 3 Emergency Stop 2 4 Warning System 2 4 Interlock System 2...

Страница 4: ...11 Chapter 7 Reconstruction of Anatomy Preview Screen 7 1 Adjusting the Panoramic View 7 3 Contour Line Setup 7 4 Removing Circumference Artifacts 7 6 Adjusting MIP Centerline and Image Type 7 6 Selec...

Страница 5: ...Folder 9 5 Reporting 9 6 Run Report 9 6 Create New Report 9 7 Modify Existing Report 9 11 Chapter 10 Calibration and Quality Assurance Panel Calibration 10 1 Geometry Calibration 10 3 QA Phantom Test...

Страница 6: ...cifications 12 1 Power Requirements 12 2 Power Cords 12 2 Environmental Specifications 12 3 Operating 12 3 Transportation and Storage 12 3 Acquisition Computer 12 3 Patient Support Chair 12 4 Disposal...

Страница 7: ...Menu A 15 Appendix B iCATVision iCATTransfer Installation iCATVision B 1 File Structure Setup B 1 Install iCATVision B 2 iCATVision Setup B 3 iCATTransfer B 4 Install iCATTransfer B 4 iCATTransfer Pa...

Страница 8: ...September 19 2007 viii Implant Planning Screen C 3 Panoramic Map C 4 Axial Slice Position C 4 Ortho Screen C 5 MPR Screen C 5 Install Case Studies from CDs C 6 TMJ Planning Screen C 6 Create Export C...

Страница 9: ...office environment The Gantry is an open design that allows Patients to sit upright during a procedure An electric powered seat is built into the Gantry for proper Patient positioning KaVo 3D eXam Ima...

Страница 10: ...amount of X ray energy delivered to the tissue CT imaging displays both high density tissue such as bone and soft tissue When interpreted by a trained Physician CT images provide useful diagnostic in...

Страница 11: ...original warranty or thirty 30 days whichever is longer The warranty period is not extended as a result of purchasing any additional parts from ISI The original purchaser must promptly notify ISI in...

Страница 12: ...ISI be liable for any special incidental or consequential damages based upon breach of warranty breach of contract negligence strict liability or any other legal theory Such damages include but are n...

Страница 13: ...d that you read this manual thoroughly including all cautions and warnings This guide uses the following conventions to describe situations that are potential hazards to the Patient or Operator potent...

Страница 14: ...al Fire Hazard Do not place any liquid or food on any part of the consoles or other modules of the system Observe all fire regulations Fire extinguishers should be provided for both electrical and non...

Страница 15: ...with large diopter or mirrors increase eye hazard with this product Radiation Safety X rays are dangerous to both operator and others in the vicinity unless established safe exposure procedures are st...

Страница 16: ...terminals 1 and 2 of the X Ray Interlock printed circuit board inside the Lower Control Center A relay contact capable of 250 volts 50 60 hertz and 3 amps is connected between these terminals When X r...

Страница 17: ...nstalled jumper between terminals 3 5 of the X Ray Interlock printed circuit board inside the Lower Control Center and connect door switches and or emergency stop switches in series between terminals...

Страница 18: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 2 6 System Labels The following labels are attached to the system...

Страница 19: ...OWER indicator lights to show that the Gantry is ON OFF removes power from the Gantry and the POWER indicators go OFF EMERGENCY STOP immediately halts all X ray and scanning activities SCAN initiates...

Страница 20: ...uct ALIGNMENT LIGHT pressing the button turns ON the laser alignment lights for three seconds Press and hold the button for three seconds turns the laser lights ON permanently until the button is agai...

Страница 21: ...s launched 3 Power up Gantry press the ON button on the Main Control Box The POWER indicator on the Main Control Box and Gantry light NOTE Although the system is now ready to scan the Ready indicator...

Страница 22: ...ame field Click OK or Overwrite an existing Workup by clicking an existing Workup Name 2 Power OFF Gantry press the OFF button on the Main Control Box The Gantry shuts down and the POWER indicators on...

Страница 23: ...ed This window lists all the Patients that were entered into the database When a Patient is selected highlighted that Patient s scanned images are listed in the window beneath the Patient List Patient...

Страница 24: ...File New Patient 2 Click Add button 3 Enter Patient data in the corresponding fields as applicable A unique Patient ID must be added such as a Patient file number or social security number Click the...

Страница 25: ...ther data has been acquired for this Patient by selecting the Done entry 4 Click OK to add Patient data 5 Click OK to close dialog box Edit Patient Details To edit Patient details 1 From the Main menu...

Страница 26: ...not scanned HIS RIS Hospital Info System Radiology Info System imported Patient Data PMS Practice Management System imported Patient Data Manual Patient Data entered manually 3 Click highlight the Pa...

Страница 27: ...log box Only Patients whose data matches the selected criteria are displayed Done Patient Data entered and scanned Queued Patient Data entered but not scanned HIS RIS Hospital Info System Radiology In...

Страница 28: ...Patient List can be hidden by selecting Tools Hide Patient from the Main menu When hidden the Patient list appears as shown below To display the Patient list select Tools Show Patient from the Main me...

Страница 29: ...ne Patient Data entered and scanned Queued Patient Data entered but not scanned HIS RIS Hospital Info System Radiology Info System imported Patient Data PMS Practice Management System imported Patient...

Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...

Страница 31: ...for Patients that are taller than 6 2 2 With the gate opened have the Patient seated in the chair facing forward The Patient should be seated comfortably with hands in lap HEAD SUPPORT BACKPLATE CHIN...

Страница 32: ...ams can cause optical damage Do not stare into the laser beam Instruct the Patient to avoid looking into the beam The use of optical instruments such as eyeglasses with large diopter or mirrors increa...

Страница 33: ...e at the chin Acquisitioning Once Patient data is entered and the Patient is properly positioned acquisitioning can begin 1 To acquire a new acquisition do one of the following Once data is entered fo...

Страница 34: ...indow is displayed 2 If desired enter any remarks in the Study Remarks text field 3 From the Quick Picks drop down list box select a standardized acquisition setup or create a new one see details belo...

Страница 35: ...cted teeth periodontal condition and other anatomy requiring detailed visualization Portrait Mode should be used for the Height of Reconstructed Volume for adult Patients where cephalometric data is r...

Страница 36: ...selected for young children and small elderly females Large is for a normal size adult or large child head size 5 Select Resolution by using the corresponding drop down list box In general the lower t...

Страница 37: ...less 6 Adjust the Volume Center slide control to select the position of the x ray sensor front to back 7 Select the Height of Reconstructed Volume from the drop down list to indicate the portrait or...

Страница 38: ...D eXam Operators Manual k990400 September 19 2007 6 8 The following dialog box is displayed 2 Click OK The system proceeds to take a Static Exposure Quick Ceph The resulting image is displayed on the...

Страница 39: ...tructed to hold still swallow before the scan take shallow breaths during the scan and may want to close his her eyes so he she won t be tempted to follow the turret with their eyes during the scan WA...

Страница 40: ...ewinds before releasing the Patient from the machine The software immediately begins the reconstruction of the image data The time required to reconstruct depends on the Scan Time and Voxel size selec...

Страница 41: ...Locator 2 Press the Alignment Light button on the Gantry Front Panel to display lasers Lasers are available while the Volume tab is selected not PAN 3 Ensure that the Vertical Laser intersects the no...

Страница 42: ...Rest and Bite Tip Holder into the Positioning Block as shown below 4 Adjust Patient height to where the Patient s chin is on the Chin Rest but keeping the Erect posture NOTE The front to back Patient...

Страница 43: ...zontal Laser Then have the Patient tilt their head downward approximately 10o Refer to the illustration above for a visual of tilting the Patient Using the Sagittal Laser front center laser ensure tha...

Страница 44: ...bserve and follow the safe exposure factors and operating instructions Do not operate this system unless you have received training to perform a procedure 12 Click the Capture button to start the test...

Страница 45: ...mber 19 2007 Acquisitioning Scanning 6 15 15 The PAN scan runs a few minutes and then displays the reconstructed image 16 Adjust the Brightness Contrast by dragging the cursor across the image vertica...

Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...

Страница 47: ...aded to the preview screen While data loading is occurring a preview window appears in the center of the screen At the bottom right side of the iCATVision window the status of the data loading is disp...

Страница 48: ...l through midline The Preview Screen is divided into seven views Patients in Database Patient Images Patient Detail View Panoramic View Sagittal View Coronal View Axial View Patients in Database lists...

Страница 49: ...Double clicking on the Coronal View brings up the MPR Screen Axial View controls the image displayed in the Panoramic View Distance and Hounsfield Unit measurements are taken on this view Double clic...

Страница 50: ...Contour Line Setup Contour Lines are used to mark the Maxilla and Mandible in the Sagittal and Coronal Views on the preview screen Maxilla is a red line Mandible is a green line Both the Maxilla and M...

Страница 51: ...ronal Views on the preview screen To do this do the following 1 On an image view on the Preview screen right click and select Add Remove Contourline 2 Click check box corresponding to the Contour Line...

Страница 52: ...em software Center Line Position click and drag the Center Line Position tool moves the displayed centerline of the selected image This changes the selected position of subordinate images as applicabl...

Страница 53: ...ark to the right of the corresponding menu item Adjusting Brightness and Contrast All views include a window level Brightness Contrast cursor tool By default this cursor is displayed When cursor appea...

Страница 54: ...creen until the cursor becomes an X 2 While the cursor is an X click the left mouse button You are returned to the preview screen This can also be performed by clicking Level Up on menu bar Filter Set...

Страница 55: ...filter setting is applied to the corresponding image Setting Filters for One or More Image Types To change the filter setting for one or more image types 1 From the Main menu select Tools Filter Sett...

Страница 56: ...d Units Shape Region for all image views These measurements calculate and display the average Mean grayscale level of the area enclosed from 1000 to 3000 where 0 equals the density of water The Standa...

Страница 57: ...tional measurements are displayed in varying colors so that they can be easily associated with the selected image area To switch off the hu cursor 1 Right click the image A popup window appears 2 Clic...

Страница 58: ...Measurements becomes available Reactivate All Measurements restores all of the inactivated measurements to the selected view Distance The system software allows making distance linear measurements for...

Страница 59: ...asurement Measurement statistics appear in the upper left corner of the image 3 Repeat steps 2 and 3 to take additional measurements A maximum of nine distance measurements can be taken at a time in a...

Страница 60: ...removed and select Remove Measurement Remove All Measurements removes all measurements from selected view Inactivate All Measurements grays out all measurement indicators and removes the measurements...

Страница 61: ...e is rotated release the mouse The grid is no longer present To complete the rotation of the entire data set 4 Right click the image and select Rotate Volume now The menu closes and the software recal...

Страница 62: ...d select Open Output Folder Microsoft Internet Explorer opens to the selected directory Saving and Loading Workups Once Patient data is acquired it is reconstructed Reconstruction allows you to make v...

Страница 63: ...in the Workup Name field c Click OK The dialog box closes and the new workup is displayed on the Select a Workup dialog box or a Overwrite an existing workup by clicking an existing Workup Name b Clic...

Страница 64: ...dialog box appears To save the changes made click Yes The Select a Workup dialog box appears Continue the save procedure as described above Load an Existing Workup To load an existing workup for a Pat...

Страница 65: ...September 19 2007 Reconstruction of Anatomy 7 19 Delete an Existing Workup To delete an existing workup 1 Right click any image on Preview screen and select Save This Workup 2 Right click the Workup t...

Страница 66: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 7 20 3 Click Yes The dialog box closes and the selected Workup is deleted 4 Click Cancel to close the Select a Workup dialog box...

Страница 67: ...ng a corresponding view on the Preview Screen Panoramic View double click displays the Implant Planning Screen Sagittal View double click displays the Ortho Ceph Screen Coronal View double click displ...

Страница 68: ...xial views which are represented on the axial slice position view and modify the criteria used to generate the axial views lower right 3D Model View lower left shows a three dimensional representation...

Страница 69: ...adjust the image displayed on the panoramic map view The blue hash mark represents the center line of the axial slices displayed on the cross section views The corresponding center cross section slice...

Страница 70: ...iew to modify the position of the cross section views which are represented on the axial slice position view and modify the criteria used to generate the cross section views at the lower right of the...

Страница 71: ...cal toolbar to adjust the slice thickness displayed on the cross section views Changes made here are reflected on the cross section views 3 Individual cross section images can be zoomed in by double c...

Страница 72: ...the Ortho screen displays a single item popup menu Clicking Tag Airways generates a 3D view of the airways for the patient in the blank view In addition the tagged airway data is displayed in the view...

Страница 73: ...airway view and select Untag Airways MPR Screen Double click Coronal View on Preview Screen displays MPR Screen The MPR Screen allows you to scroll through the Axial upper left Sagittal upper right an...

Страница 74: ...ical toolbars to adjust the slice thickness of the corresponding color coded view To view the result of a linear slice 1 To view a linear slice right click the view and select Line A pointer cursor is...

Страница 75: ...sulting slice is displayed in the view at the lower right of the MPR Screen 3 To reposition the line drag the end points of the line To explore additional cut planes in an animated consecutive fashion...

Страница 76: ...to position the cut planes slice as desired Cursor changes to a film icon 6 Right click image and select Explore Speed 7 Enter desired animation speed in Movie Playback Speed field 8 Click OK 9 Click...

Страница 77: ...TMJ Screen TMJ Screen enables condyle mapping and creating corresponding coronal slice views To use the TMJ Screen 1 If necessary pan the Axial SMV View upper left down in the window to see the condyl...

Страница 78: ...lbars to adjust the slice thickness of the cross section views To create lateral slices 5 Drag the center blue circles on the Axial View to move the corresponding Condyle Map 6 Drag the yellow and blu...

Страница 79: ...s This procedure is used to determine where these DICOM image files are to reside 1 From the top Main menu select Tools Setup DICOM Database Root Folder contains patient data that is required for crea...

Страница 80: ...ot presently used Check Read Write Access to Image Database This function tests the ability of iCATVision to read write to the image root folder iCATVision requires Write Access to this path to store...

Страница 81: ...tandard for digital images which is a means for medical imaging products to share images The following process demonstrates how to convert patient data files into DICOM images for export To Export DIC...

Страница 82: ...ntent the new files are stored to the folder replacing old files The Destination field was setup at the start of this section 4 Click OK Done is displayed when the process is complete Create CD The in...

Страница 83: ...rase CD RW This Erase CD option can only be used if using re writable CDs 1 From the top Main menu select Tools Create Export CD The CD Burner window is displayed shown above 2 Click Erase CD RW An op...

Страница 84: ...ent information and text notes can also be inserted as required Once these reports are created and saved they are readily available for printing and distribution Patient reports can also be accessed f...

Страница 85: ...st Allow time for the Patient Image to load 2 From the Main menu select Tools Reporting 3 Click Create New Report from the Options box 4 The Edit Fields box is displayed Enter all required Clinic and...

Страница 86: ...Life Size used to depict the image in life size able to measure directly on the paper The drop down list box allows you to specify where the image originated For example if Left Top Corner is selecte...

Страница 87: ...e image Insert Text Box 8 Select Insert Text from the Main menu or right click the report page and select Add Text 9 Select desired variable from the Change Variable list Holding down the CTRL key and...

Страница 88: ...mouse button and dragging the cursor Cross Arrows used to move Text box Double Headed Arrow used to size Text box Add Page 13 Select Insert Page from the Main menu or right click a report page and sel...

Страница 89: ...eport This function is used to edit reports that were previously created To create a new Patient Report 1 From the Main menu select Tools Reporting 2 Click Modify Existing Report from the Options box...

Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...

Страница 91: ...Landscape positions The entire calibration takes about 8 to 10 minutes to complete The process runs a five mode calibration as listed below Mode 0 4 x 4 Landscape Mode 1 2 x 2 Landscape Mode 2 2 x 2 P...

Страница 92: ...m the field of view and click OK to start X ray exposure 7 Click OK 8 The Panel Calibration runs for about 10 minutes The progress of the scan is displayed on the Progress bar at the bottom of the Acq...

Страница 93: ...mplete 11 Perform the Panel Calibration procedure listed above 12 Mount the Phantom Platform shown below 13 Ensure that the BB Phantom is level 14 Center the BB Phantom on the platform using the align...

Страница 94: ...f required make adjustments and click Preview again Repeat as required To move the phantom to the right or left on the Preview screen use the Back Front feature on the Volume tab and then return to th...

Страница 95: ...bove the green progress bar ensure the following Landscape Scan 24 Beads detected 24 Beads valid Portrait Scan 20 Beads detected at least 16 Beads valid 24 Click OK when the calibration completes NOTE...

Страница 96: ...Phantom Test and QA Water Test This procedure is performed to check the High Contrast Spatial Resolution 1 Remove Chin Cup and insert Phantom Platform 2 Place QA Phantom on Platform Use a piece of foa...

Страница 97: ...ck Add or right click an existing image from database and select Acquire Now If adding a new patient enter QA Test in the First Name field and Line Pair in the Patient ID field 7 Click OK to close the...

Страница 98: ...achieve the proper height 13 To move the phantom to the right or left use the Back Front feature on the Volume tab and then return to the Calibration tab If required make adjustments and click Preview...

Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...

Страница 100: ...iew bottom image 3rd from the left 19 The image below is displayed In the upper right corner view slide the Vertical and Horizontal lines to the Line Pairs centers as shown below 20 Zoom Line Pair ima...

Страница 101: ...sent within line pairs 10 11 and 12 Compare your image quality to the image shown above Your Line Pair 12 should have the same image quality definition as the 13 Line Pair displayed above or better Di...

Страница 102: ...pper left To zoom start at the lower right corner of the image and drag the cursor across the image For consistency ensure that the slice selection to be measured is setup as follows a Right click top...

Страница 103: ...ed within the QA phantom Note to achieve an accurate measurement boxes should be as large as possible without touching the circle diameter 31 Note and record the Hounsfield value of each material See...

Страница 104: ...of rotation and the horizontal laser line is located across the center of water depth 5 Start test by selecting File New Patient from the Main menu 6 From the Patient Information Panel click Add or r...

Страница 105: ...f the water height Adjust the Phantom Platform to achieve the proper height Use the menu Back Front arrow slide to adjust the phantom side to side position If required make adjustments and click Previ...

Страница 106: ...Operators Manual k990400 September 19 2007 10 16 13 The red dotted line in the upper right screen should appear in the middle of the water height Use the Center Line Positioning tool to make the heigh...

Страница 107: ...terest in the center of Water and also the Air outside area of the water bowl as shown above HU Area box size should be approximately 1000 0 mm2 inside the Water Phantom area 400 0 mm2 outside the Wat...

Страница 108: ...e Level Test right click the HU data upper right corner of image and select Remove all measurements 19 Right click the image and select HU Statistics from the menu Use the Region Tool to define a ROI...

Страница 109: ...ording values a fifth ROI is required from the center of the water area The fifth measurement replaces the first measurement since four measurements is the limit 22 Note and record the Mean and SD val...

Страница 110: ...alls facing up 4 Insert Bite Tip into the phantom front mounting hole to lock phantom into place 5 Press the Alignment Light button on the Gantry Front Panel to display lasers Use the Horizontal Laser...

Страница 111: ...e Patient Information box 9 Click the Pan tab in the Acquisition window and select Exposure Large 10 Click the Capture button to start the test 11 The system moves approximately 1 4 rotation to the Ho...

Страница 112: ...runs a few minutes and then displays the reconstructed image 14 Adjust the Brightness Contrast by dragging the cursor across the image vertical horizontal All seven metal balls should become visible...

Страница 113: ...axial lasers intersect This is in the lower center of the detector 2 Perform a 9 second 13 cm 0 4 voxel scan and record time and dose from meter 3 Repeat step 2 for 27 seconds and record Measured Dose...

Страница 114: ...presentative Imaging Sciences International LLC Guy Long 1910 North Penn Road Imaging Sciences International LLC Hatfield PA 19440 Western Europe Service Centre United States of America Corinium Indus...

Страница 115: ...JIS T0601 CE MDD 93 42 EEC IEC EN 60601 1 LVFS 2003 11 Swedish regulation transposing the MDD 93 42 EEC IEC EN 60601 1 1 CMDCAS Canadian Medical Device Regulation IEC EN 60601 1 2 ISO 10993 1 2003 IEC...

Страница 116: ...Canada Medical Device Keyword Index for X ray equipment and IEC standards 60601 2 7 and 60601 2 28 The following class of equipment was determined Protection against electric shock Class I Applied pa...

Страница 117: ...rious bodily injuries to Patients and persons in the surrounding area Adequate precautions must always be taken to avoid or reduce exposure to the useful beam as well as scattered radiation Refer to t...

Страница 118: ...f 41 7 mA min week 50 scans per week yielding a workload of 83 3 mA min week Total scan time 9 seconds Location Distance in Feet meter Exposure mR Exposure R 10 scans wk mR wk 25 scans wk mR wk 50 sca...

Страница 119: ...11 110 1 10 2 75 5 50 9 2 74m 0 036 36 0 36 0 90 1 80 135o 3 0 91m 0 031 31 0 31 0 78 1 55 6 1 82m 0 021 21 0 21 0 53 1 05 9 2 74m 0 012 12 0 12 0 30 0 60 180o C 3 0 91m 0 124 124 1 24 3 10 6 20 6 1...

Страница 120: ...a minimum of two meters or behind a properly shielded permanent barrier A viewing window or alternative method such as a mounted mirror should be present to enable the X ray Operator to view the Patie...

Страница 121: ...ead Phantom and used a light localizer to position phantom in center of scan so slice cuts through the center of the phantom 2 Performed a Capture view to assess the alignment 3 Performed a scan 4 Use...

Страница 122: ...hicknesses is not required as this is not applicable to the 17 19 Cone Beam Volumetric Tomography technology 10 Repeat for chamber in peripheral locations 1 cm interior to the surface of the phantom w...

Страница 123: ...11 7 Sensitivity Profile For the ultra high resolution mode when the bead SIZE in the phantom is scanned the following sensitivity profile is obtained mm Hounsfield value 1 2752 2 3104 3 2560 4 3104...

Страница 124: ...semblies Operator Data Available focal spot sizes mm 0 5 Available target angles 15 Anode construction Vacuum cast copper with tungsten target Cathode construction Vacuum tube nickel with tungsten fil...

Страница 125: ...ied Sige Phase Bi polar FILAMENT CURRENT AMP TUBE CURRENT mA FILAMENT VOLT AMP HARACTERISTICS 60 Hz AC DC VOLTS AMPS ANOD OOLING CURVE ANODE HEAT STORAGE AACITY HU X 1000 TIME IN MINTS H U Heat Unit k...

Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...

Страница 127: ...cations in order to have the focal spot to skin distance as large as possible Minimum Filtration at 120 kVp mm of aluminum equivalent 10 mm or greater Maximum Rated Continuous Tube Operation 130 kVp 0...

Страница 128: ...x 6 67 inches height maximum 23 2 cm diameter x 17 cm height maximum Note Maximum values can be collimated down Primary Reconstruction Less than 2 minutes for 27 sec scan 0 4 voxel Secondary Reconstr...

Страница 129: ...torage 32 to 104 degrees Fahrenheit 0 to 40 degrees Celsius 30 to 80 humidity Acquisition Computer Minimum Specifications for Acquisition Computer Provided with the System Pentium 4 Processor 3 8 GHz...

Страница 130: ...X ray source assembly image sensor and all electronic circuits should be regarded as non environmental friendly waste product Extension Cords Do not use any extension cords which are not provided with...

Страница 131: ...onment Guidance IEC 61000 4 2 Electrostatic Discharge ESD 6 kV Contact 8 kV Air Floors should be wood concrete or ceramic tile If floors are synthetic the relative humidity should be at least 30 EFT I...

Страница 132: ...tions equipment Maximum Output Power Watts Separation 150 kHz to 800 MHz Separation 800 MHz to 2 5 GHz 0 01 15 in 38 cm 30 in 76 cm 0 1 15 in 38 cm 30 in 76 cm 1 46 in 117 cm 8 ft 2 4 m 10 12 ft 3 7 m...

Страница 133: ...EN 60601 1 4 CSA C22 2 No 601 1 IEC EN 60601 2 32 JIS Z4701 IEC EN 60601 2 44 JIS Z4703 IEC EN 60825 1JIS T0601 JIS T0601 CE MDD 93 42 EEC IEC EN 60601 1 LVFS 2003 11 Swedish regulation transposing t...

Страница 134: ...PONSIBILITY IN THOSE CASES WHERE NON COMPLIANCE WITH THE STANDARD RESULTS FROM THE USER S FAILURE TO HAVE THE MANUFACTURER S RECOMMENDED MAINTENANCE PERFORMED The actual maintenance inspection and con...

Страница 135: ...ification Label X X Warning and Indicators X X Oil Leaks X X Physical Damage X X Mounting System Stability X X Beam Limiting Device Physical Damage X X Certification Label X X X Ray Controller Timer X...

Страница 136: ...r 200 Volt Systems K1304 0 Patient E stop Box K1305 Patient Alignment Panel K1306 0 Operator Control Box 1539 Laser Cross 1540 Laser Straight Line 1543 Motor Beam Limiter 7 5 Stepper K1701 0 Isolation...

Страница 137: ...1 10 Glide 1000179 4 11 Control Box 1306 0 1 12 Chair Patch Cable 1520 1 13 Cable Clips 101 6 6 14 Head Rest 27 0 1 15 Head Band 27 1 100 16 Pan Phantom 12 0 1 17 Phantom Assembly 14 1 0 1 18 Line Pa...

Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...

Страница 139: ...Three Dimensional Volume Rendering 3DVR Open Database 3 DVR is a standalone program which runs independently and is not part of iCATVision 1 To open DVR Diagnostic Volume Rendering double click the DV...

Страница 140: ...n DICOM format If importing a case from iCATVision make sure to export the case in DICOM3 Multi File format 2 Click Open Dataset a browse window is displayed 3 Select the drive folder for the DICOM da...

Страница 141: ...d DVR In this sample it is slice 10 of 256 To scroll through each slice 1 Click Paging Radio button 2 Drag the cursor either UP or DOWN to scroll through the axial slices Dragging the cursor upward sc...

Страница 142: ...vate the ROI tool 1 Click ROI Radio button 2 Drag cursor to create a box around region of interest Release the mouse button to move box then click again to complete the box Calculated items are displa...

Страница 143: ...distance tool measures linear distance To activate Distance tool 1 Click Distance radio button 2 Drag cursor from point to point to create a measurement line see below in green Linear measurement is...

Страница 144: ...e Patient Chin Cup that may appear in the scan See the sample below Patient Chin Cup highlighted in blue To activate Identify tool 1 Click Identify Box The 3DVR Setup window is displayed and the dense...

Страница 145: ...er in the box Remove Object To remove a piece of anatomy from image 1 Click the Remove Object box to enable checkmark 2 Then click the object of interest highlighted in blue 3 Close 3DVR Setup window...

Страница 146: ...k for a region that has relatively even grayscales 3 At this location enable the ROI function 4 Drag the cursor in the area between the dental arch making sure not to include any teeth or bone This di...

Страница 147: ...n Offset which currently has a value of 0 7 Click OK View Volume View Volume allows viewing all 3 projections of the 3D data Axial Coronal and Sagittal Also enables the creation of 3D Renderings Enabl...

Страница 148: ...he Axial Coronal and Sagittal views are all in Radiographic mode NOTE Window Level can be adjusted in each of the 3 views by dragging the cursor in each window MIP mode Maximum Intensity Projection vi...

Страница 149: ...of interest in the Coronal and Sagittal views Again the selected data is used to calculate the 3D image Once the area of interest is determined the 3D image can be displayed by utilizing the VR Funct...

Страница 150: ...te a 3D Image 1 From the first View drop down menu select a direction to display The choices are AP Anterior Posterior PA Posterior Anterior LL Left Lateral RL Right Lateral Up Down 2 From the second...

Страница 151: ...ample the entire volume data is selected in all 3 views Below is a sample of a 3D image calculated from the selected volume area seen below Sample of View Types Each View Type has its own Parameters w...

Страница 152: ...D image to Full Screen Right click on the 3D image and select View Full Screen PAN Move the 3D image within the window by dragging cursor on image Zoom Zoom 3D image Drag cursor on image UP to Zoom OU...

Страница 153: ...age in current working folder adds jpg extension The current working folder is the folder opened from the Open Dataset Save Grayscale to TIFF Saves 3D image in current working folder adds tiff extensi...

Страница 154: ...entire computer screen All the same manipulation functions and 3D options are used in this display Right click to display Pop Up menu to utilize functions To return to the original viewing format pres...

Страница 155: ...limit on usage and the number of copies File Structure Setup 1 On the iCAT Acquisition Computer create an iCATVision folder under the C drive 2 Create a subfolder under iCATVision called iCATVision Wa...

Страница 156: ...AT Acquisition Computer s C drive under the iCAT Vision Folder Create a new Folder 1 Open the folder where the new folder is to reside 2 Right click and select New Folder 3 Type the name for the new f...

Страница 157: ...the C drive or on a networked drive Setup is the same on all networked and non networked computers or workstations 3 A prompt to restart iCATVision software is displayed click OK NOTE It is currently...

Страница 158: ...der The Image Root folder typically resides on a server or another central storage location If the computers are on the same network as the iCATTransfer then there is direct access to the patient imag...

Страница 159: ...Agreement and then follow the instructions below iCATTransfer Setup The program must be configured to watch for exported DICOM data from iCAT in order to authenticate and transfer it to the image root...

Страница 160: ...3 Under Image Root Folder Primary Destination Browse to iCATVision Root folder and select This is most likely on a server central storage location 4 Click Start This saves the preset Watch and Root d...

Страница 161: ...k990400 September 19 2007 B 7...

Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...

Страница 163: ...k990400 September 19 2007 B 9...

Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...

Страница 165: ...ocation 1 2 3 5 6 7 4 REMINDER Images are displayed as if you are looking at the patient from the front TO DISPLAY PATIENT IMAGE 1 Click Patient Name 2 Click Patient Scans DISPLAYED VIEWS 3 Patient St...

Страница 166: ...Filter Settings Reset to Default Removing Circumference Artifact Circumference Artifact are seen visually in the Preview Screens as horizontal lines in the Coronal and Sagittal images and a white part...

Страница 167: ...LS All views except the 3D Model have Brightness Contrast Rotate Drag Zoom and Pan 3D Model only has Rotate Back Tool to exit out of a planning screen back to the Main Display move cursor to the very...

Страница 168: ...this tool to adjust slice thickness of Panoramic View Drag the center tool to adjust Pan Focal Trough Click the bottom tool to change the Pan view from Radiographic to MIP 3 Vertical Tool Bar Drag thi...

Страница 169: ...ank area 1 3 1 2 REMINDER Ortho Screen is acquired by double clicking Sagittal View from Preview window DISPLAYED VIEWS The Ortho Screen displays the Lateral Cephs in Radiographic and MIP mode as well...

Страница 170: ...to be loaded Just click the patient name TMJ Planning Screen 1 Drag center tool to scroll up and down Sagittal view to locate condyles in the Axial view so you can see condyles properly for mapping 2...

Отзывы: