k990400 September 19, 2007
Calibration and Quality Assurance
10-17
Noise Level Test
14. Zoom the Axial image (upper left). To zoom, start at the lower
right corner of the image and drag the cursor across the image.
15. Right click the image and select
HU Statistics
from the menu.
Use the Region Tool to define a ROI (Region of Interest) in the
center of Water and also the Air (outside area of the water bowl),
as shown above.
HU Area (box size) should be approximately:
1000.0 mm
2
inside the Water Phantom area
400.0 mm
2
outside the Water Phantom area (AIR)
16. Note and record the values of both the Water and Air areas. See
chart below.
NOTE:
Do not close image. This image is also used for the
Uniformity Test (next procedure).
17. If the Standard Deviation (
SD
) Measured Value
exceeds 100
,
call Service.
Measured Values
Water
Air
Mean
Expected Values
0 +50
-1000 +50
SD
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...