Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
2-4
Emergency Stop
This manual contains instructions for safe operation of this dental X-
ray System. In the event of an emergency (any moving component
collides with any parts of the equipment or items in the environment,
or that could cause physical injury to the Patient), the Operator
should utilize the
Emergency Stop button
to turn off the power to
all moving parts in order for the Patient to be safely removed from
the machine.
Warning System
The System is equipped with provisions for warning lights and/or
audible alarms when X-ray power is energized. There are two
methods available to use this system:
Method 1: Externally provided power
An externally powered warning system can be connected to
terminals 1 and 2 of the X-Ray Interlock printed circuit board inside
the Lower Control Center. A relay contact capable of 250 volts, 50/
60 hertz, and 3 amps is connected between these terminals. When X-
ray power is energized this relay contact is closed and the warning
system energized.
Method 2: Equipment provided power
A warning system may be powered from 12 volts DC within the
System. The warning system cannot exceed 12 volts DC, 3 Amps.
To use this provision connect the warning system as follows:
1. Connect the “common” of the warning system to terminal 6 of
the X-Ray Interlock printed circuit board inside the Lower
Control Center.
2. Connect terminal 1 to terminal 5 on the circuit board.
3. Connect the “positive” of the warning system to terminal 2 of the
circuit board.
When X-ray power is energized 12 volts DC will be provided
through the relay contact between terminal 1 and 2 of the X-ray
Interlock printed circuit board and energize the warning system.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...