k990400 September 19, 2007
10-1
Chapter
10
Calibration and Quality
Assurance
The Panel and Geometry Calibrations can be conducted by the
Owner / Operator of the device. It is recommended that the Panel
Calibration be performed once a week.
Panel Calibration
The Panel Calibration runs in both Portrait and Landscape positions.
The entire calibration takes about 8 to 10 minutes to complete. The
process runs a five mode calibration as listed below:
•
Mode 0 (4 x 4) Landscape
•
Mode 1 (2 x 2) Landscape
•
Mode 2 (2 x 2) PAN (Landscape)
•
Mode 3 (4 x 4) Portrait
•
Mode 4 (2 x 2) Portrait
To run the Panel Calibration:
1. Ensure room temperature is in the range of 50 to 95
o
F (10 to
35
o
C).
2. From the Main menu select
File > New Patient
.
3. From the Patient Information Panel, click
Add
or
open (double click) an existing
Panel Calibration
scan.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...