Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
10-2
4. Click
OK
to close Patient Information box.
5. Click the
Calibration
tab.
6. Click the
Panel Calibration
button (bottom button).
A window is displayed
“Remove all objects from the field of
view and click OK to start X-ray exposure.”
7. Click
OK
.
8. The Panel Calibration runs for about 10 minutes. The progress
of the scan is displayed on the
Progress
bar at the bottom of the
Acquire window.
”Panel Calibration is complete”
is displayed when the process
is finished.
9. Click
OK
.
10. If you are running the Geometry Calibration next, then do not
close the Calibration screen when the Panel Calibration is
completed.