22. The Geometry Calibration scans both Landscape and Portrait
Modes.
NOTE:
Ensure that the metal platform support is below the Field of
View (as shown above.)
23. Calibration progress is displayed on the bottom of screen.
Ensure that all beads are detected and valid in both scans. Data
is displayed above the green progress bar, ensure the following:
•
Landscape Scan -
24 Beads detected, 24 Beads valid
•
Portrait Scan -
20 Beads detected, at least 16 Beads valid
24. Click
OK
when the calibration completes.
NOTE:
If the calibration did not detect a bead at a certain location,
check to ensure that all beads are in tact on the BB Phantom.
25. Restart iCATVision software. This completes the Geometry
Calibration.
CAUTION
CAUTION
Failure to restart iCATVision software at this point may result in a
system failure.
LANDSCAPE SCAN
PORTRAIT SCAN
Metal Platform Support
(below FOV)
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...