Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
12-6
If the System does cause harmful interference to surrounding
equipment which can be determined by turning the system OFF and
ON, the customer or user is encouraged to try to correct the
interference by one or more of the following methods:
•
Increase separation between the system and surrounding
equipment.
•
Connect the system into an outlet on a circuit different from
that to which the surrounding equipment is connected.
Radiated RF
IEC 61000-4-3
The 17-19 Imaging System is intended for use in an
electromagnetic environment in which radiated
disturbances are controlled. The customer or user
of this system can help prevent electromagnetic
interference by maintaining a minimum distance
between portable and mobile RF communications
equipment and the system as recommended
below, according to the maximum output power of
the communications equipment.
Maximum
Output Power
(Watts)
Separation
150 kHz to 800
MHz
Separation
800 MHz to 2.5
GHz
0.01
15 in (38 cm)
30 in (76 cm)
0.1
15 in (38 cm)
30 in (76 cm)
1
46 in (117 cm)
8 ft (2.4 m)
10
12 ft (3.7 m)
24 ft (7.3 m)
100
38 ft (11.7 m)
77 ft (23.4 m)
Note:
Field strengths from transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level of 3 V/m.
Emission Test
Compliance
Electrostatic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The system uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The system is suitable for use in all establishments,
including domestic, and those directly connected to
the public low - voltage power supply network that
supplies buildings for domestic purposes.
Harmonics
IEC 61000-3-2
Class A
Flicker
IEC 61000-3-3
Complies
The System is intended for use in an electromagnetic environment specified below. The
Customer or User should ensure that the system is used in such an environment.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...