k990400 September 19, 2007
Radiation Environment Survey
11-5
•
An annual radiation survey may be typically required. This
survey is typically required to be submitted to the local
agency or government body.
•
A phantom or patients may be used for system training.
Employees of the facility may not be used for this training.
•
The system shall be registered with the local agency or
government body.
Dose and Imaging Performance Information
Conditions of Operation
The dose measurements were made using a PMMA Head Phantom
with a diameter of 16 centimeters and a 10cc ion chamber with an
active length of 10 centimeters.
Procedure Performed:
1. Leveled the Head Phantom and used a light localizer to position
phantom in center of scan, so slice cuts through the center of the
phantom.
2. Performed a Capture view to assess the alignment.
3. Performed a scan.
4. Used image viewer to verify that the center hole is within 5 mm
from the center of the scan in all directions. The central position
is a position where the CTDI values are the maximum.
5. Placed the ionization chamber in the center hole. Inserted acrylic
rods in all other holes.
6. Used typical technique factors.
7. Performed a single scan and read out of the ionization detected
with the CT chamber X, in coulombs (C).
8. Calculated CTDI using the equation:
CTDI (rad) = Xr x
f
rad/R x 10(cm) / NST (cm)
Where:
•
Xr = the measured exposure in Roentgens
•
f
= a factor to convert exposure in air to absorbed dose in
tissue or other attenuating matter. One must use the f-factor
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...