PAN Phantom Test
PAN Phantom Test is used to validate the PAN scan data capture.
To perform a PAN Phantom Test:
1. Prepare the Bite Tip by inserting the narrow edges of the Bite
Tip down into the Bite Tip Holder uprights. Then turn the Bite
Tip a 1/4 turn to lock into place.
2. Insert the Chin Rest and Bite Tip Holder into the Positioning
Block.
3. Slide PAN Phantom into Chin Rest slot (balls facing up).
4. Insert Bite Tip into the phantom front mounting hole to lock
phantom into place.
5. Press the Alignment Light button on the Gantry Front Panel to
display lasers. Use the Horizontal Laser to adjust the height of
the phantom as shown below.
Lasers are available while the Volume tab is selected (not PAN).
6. Start test by selecting
File > New Patient
from the Main menu.
7. From the Patient Information Panel, click
Add
or right click an
existing image from database and select
Acquire Now
.
If adding a new patient, enter
PAN Test
in the
First Name
field
and
QA
in the
Patient ID
field.
PAN
PHANTOM
CHIN REST
BITE TIP HOLDER
BITE TIP
LASER
LINE
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...