k990400 September 19, 2007
Acquisitioning (Scanning)
6-5
•
2 arches/anchor, 8.5 sec, 0.3 Voxel -
recommended when
data is exported to implant planning software systems such
as Nobel or SimPlant.
Please check with your service to determine the desired
height of the scan. Please be aware that most 3rd party
software currently requires the scan to be taken at 0.4 Voxel
resolution and no smaller.
•
Mandible or Maxilla…
-
Protocols can be used if only 1
arch is required for an Implant case.
•
High Resolution (0.2, 0.25, 26 sec.) -
recommended
protocols for specialized cases such as 3rd molars, root
canal relationships, small root fractures, supernumerary or
impacted teeth, periodontal condition and other anatomy
requiring detailed visualization.
•
Portrait Mode
-
should be used for the Height of
Reconstructed Volume for adult Patients where
cephalometric data is required.
2. If the standardized setups do not meet your exact requirements,
either select all of the desired criteria for the scan
or
Select a
Quick Pick
and make modifications to the available
criteria.
To add new criteria to the Quick Picks list:
a. Click the
Save…
button, below the Quick Picks list box.
The following dialog box appears.
b. Enter a name for the new Quick Pick in the text field on the
dialog box.
c. Click
OK
. The dialog box closes. The newly created Quick
Pick is now available from the Quick Picks drop-down list
box.
To delete a Quick Pick from the list:
a. Select the Quick Pick to delete from the Quick Picks drop-
down list box.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...