k990400 September 19, 2007
Acquisitioning (Scanning)
6-3
7. Make sure the Patient’s chin is stabilized in the chin support.
8. Slide the head support forward for additional head stabilization.
9. Secure the head strap around the forehead.
10. For scans that are specific to TMJ or orthodontics, only the head
support and head strap should be used. It is recommended that
the chin support not be used because it can affect the Condular
position in Fossa and interfere with soft tissue at the chin.
Acquisitioning
Once Patient data is entered and the Patient is properly positioned,
acquisitioning can begin.
1. To acquire a new acquisition do one of the following:
Once data is entered for the Patient in the Select Patient dialog
box, with the Patient name selected (highlighted) click the
OK
button.
- or -
From the Patient Information Pane on the preview screen,
double-click the Patient for whom new scans are to be acquired.
Tilt:
Occlusal Plane Flat,
Frankfur Plane tilted
up slightly.
Position of Alignment Lights for Full Head Scan:
Horizontal:
Occlusal plane between the lips.
Vertical:
1.5 inches in front of condyle (chin support adjusted to CT acquisition position.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...