k990400 September 19, 2007
Calibration and Quality Assurance
10-7
4. Center the QA Phantom on the platform with the Air Hole
positioned at the left rear of the Gantry (shown below).
5. Start the test by selecting
File > New Patient
from the Main
menu.
6. From the Patient Information Panel, click
Add
or right click an
existing image from database and select
Acquire Now
.
If adding a new patient, enter
QA Test
in the
First Name
field
and
Line Pair
in the
Patient ID
field.
7. Click
OK
to close the Patient Information box.
8. From the Volume tab in the Acquisition window, select the
following:
Exposure:
Large
Height of Reconstruction Volume:
Landscape 6cm
Resolution: .2 voxel
9. Click
Preview
to check the phantom position.
10. Click
OK
on the settings screen.
AIR HOLE
ACRYLIC
TEFLON
LDPE
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...