Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
10-18
Uniformity Test
18. Use the Noise Level Test dataset (previous procedure), if not
available, perform steps 1 to 13.
If using the Noise Level Test, right click the HU data (upper
right corner of image) and select
Remove all measurements
.
19. Right click the image and select
HU Statistics
from the menu.
Use the Region Tool to define a ROI (Region of Interest) in four
areas of Water, as shown above.
NOTE:
Four regions is the number of regions that can be displayed
at one time. Creating another region (fifth) will overwrite the first.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...