Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
10-6
QA Phantom Test
The following Quality Assurance Tests can be conducted by the
Owner / Operator of the device. It is recommended that the System
Quality Assurance be performed annually or if image quality
becomes degraded. For this purpose, the following procedures are
provided with a QA Phantom Test and QA Water Test.
This procedure is performed to check the High Contrast Spatial
Resolution.
1. Remove Chin Cup and insert Phantom Platform.
2. Place QA Phantom on Platform. Use a piece of foam beneath the
phantom to elevated it above the Metal Platform Support Arm,
as shown below.
3. Use the alignment lasers to adjust the Phantom Platform height.
Adjust the height so that the Axial Laser is positioned at the
center of the QA Phantom.
Horizontal Laser Line
through Center of Phantom
Metal Platform Support Arm
is Below Bottom of Phantom
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...