An AM delay measurement characterizes the group delay (or envelope delay)
of a device. lb perform this measurement you must have ordered either
Option
(AM Delay, 50 ohm) or Option
(AM Delay, 75 ohm). These
options include internal instrument hardware and
two external
scalar detectors and a power splitter.
Group delay flatness can be a key specification for many components and
systems. Distortionless transmission of a signal requires constant amplitude
and group delay response over the frequency bandwidth. Group delay is the
measurement of signal transmission time through a device. It is defined as the
derivative of the phase characteristic with respect to frequency.
If the device under test is a frequency translator, the device input and
output frequencies will by
be different. This generally makes the
measurement of the device phase response (and therefore group delay) very
The AM Delay option overcomes this
by using an amplitude
modulation technique to measure group delay. In this technique, a small
amount of amplitude modulation is applied to the RF output of the analyzer.
Scalar detectors are used to detect this modulation both before and after
the device under test. The group delay can then be calculated from the
phase difference between these two signals (modulation envelopes). Since
broadband detection is used, the Option
analyzer can measure delay
through nearly any device, including frequency translators.
There are several important considerations in an AM delay measurement
the device is a limiter or has AGC (automatic gain control), this will tend to
distort or remove the amplitude modulation used for the measurement. Any
limiting or AGC in the device should be disabled before making an AM delay
measurement. The broadband detection used for AM delay is susceptible to
spurious signals and noise. High-level spurious signals should be removed
with
The signal levels at both the reference and test detectors
should be kept as high as possible. The specified incident power range for
both detectors in an AM delay measurement is 10 to + 13
If the
device
power must be outside this range,
or attenuation
must be used directly before the device. If the device
power is outside
the 10 to + 13
range, attenuation or amplification must be used directly
after the device.
3-42
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Содержание 8711C
Страница 18: ...I 1 Installing the Analyzer 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 103: ...4 Using Instrument Functions 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 203: ...5 Optimizing Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 258: ...7 Automating Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 452: ...I 10 Specifications and Characteristics 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt co ...
Страница 473: ...11 Safety and Regulatory Information 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 480: ...12 Preset State and Memory Allocation 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...