Calibrating for Increased Measurement
Accuracy
Check the Calibration
‘lb perform a calibration check:
1. Be sure that the analyzer is in the measurement mode and has
implemented the particular calibration that you want to check.
Press
Da Cal
.
3. Follow the
prompts to connect the calibration standards.
(You should connect the standards at your calibration reference plane,
not necessarily at the front panel of the analyzer. For example, your
calibration reference plane may be at the end of cables, adapters, or test
fixtures connected to the analyzer. See “The Calibration Reference Plane”
for more information
C A U T I O N
To decrease the likelihood that a faulty calibration standard is being used,
it is recommended that you use a
set of calibration standards for
the calibration check than you used to perform the initial calibration. The
different set of calibration standards must be of the same type as the original
4. Once you have properly connected all the required standards as prompted,
the analyzer computes the residual errors for the measurement.
Now press the
View
key to view the residual errors for the
type of calibration you did a calibration check on. When you view an
error term plot, the analyzer stops sweeping and displays the plot on the
analyzer’s screen. lb return to measurement mode and autoscale the
display, press
.
6. Refer to “Error Term Descriptions and Typical Values” for descriptions and
typical plots of the various error terms.
The following table lists the typical error term values.
6-28
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Содержание 8711C
Страница 18: ...I 1 Installing the Analyzer 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 103: ...4 Using Instrument Functions 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 203: ...5 Optimizing Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 258: ...7 Automating Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 452: ...I 10 Specifications and Characteristics 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt co ...
Страница 473: ...11 Safety and Regulatory Information 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 480: ...12 Preset State and Memory Allocation 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...