I
Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
There are several situations that may require you to
your own
calibration kit
Here are three examples:
l
You are using a connector type (TNC or BNC for example) which is not one
of the selections under the CAL KIT menu.
l
You are using a built-in connector type, but your standards (such as a short)
have different characteristics than the built-in kit uses when performing a
calibration.
You are using a test
The following steps (explained later in detail) should be performed when
defining your own cal kit:
1. Determine the standard characteristics for the connector type you plan to
use.
2. Create a
kit ASCII file or edit the one provided on the Example
Programs Disk.
N O T E
Calibration kit definitions
be in DOS format. format is not supported for cal kit definitions.
3. Verify performance.
Step 1: Determine the
Determine the standard characteristics for the connector type you plan to use.
Standard
In particular, these are the characteristic impedance
the delay, and the
Characteristics
loss. These characteristics are common for the four supported types used by
the analyzer. Additionally, the capacitive model parameters, CO, Cl,
and
are necessary for fully
the open. These electrical characteristics
can be mathematically derived from the physical dimensions and material of
each calibration standard or from its actual measured response.
6-19
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Содержание 8711C
Страница 18: ...I 1 Installing the Analyzer 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 103: ...4 Using Instrument Functions 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 203: ...5 Optimizing Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 258: ...7 Automating Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 452: ...I 10 Specifications and Characteristics 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt co ...
Страница 473: ...11 Safety and Regulatory Information 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 480: ...12 Preset State and Memory Allocation 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...